A Performance Comparison of Nine Selected Liquid Scintillation Cocktails

Research output: Report/bookBLG - Open report

Standard

A Performance Comparison of Nine Selected Liquid Scintillation Cocktails. / Verrezen, Freddy; Loots, Hilde; Hurtgen, Christian.

0 ed. SCK CEN - Belgian Nuclear Research Center, 2008. 132 p. (SCK•CEN Reports; No. BLG-1052).

Research output: Report/bookBLG - Open report

Harvard

Verrezen, F, Loots, H & Hurtgen, C 2008, A Performance Comparison of Nine Selected Liquid Scintillation Cocktails. SCK•CEN Reports, no. BLG-1052, vol. 1, 0 edn, SCK CEN - Belgian Nuclear Research Center.

APA

Verrezen, F., Loots, H., & Hurtgen, C. (2008). A Performance Comparison of Nine Selected Liquid Scintillation Cocktails. (0 ed.) (SCK•CEN Reports; No. BLG-1052). SCK CEN - Belgian Nuclear Research Center.

Vancouver

Verrezen F, Loots H, Hurtgen C. A Performance Comparison of Nine Selected Liquid Scintillation Cocktails. 0 ed. SCK CEN - Belgian Nuclear Research Center, 2008. 132 p. (SCK•CEN Reports; BLG-1052).

Author

Verrezen, Freddy ; Loots, Hilde ; Hurtgen, Christian. / A Performance Comparison of Nine Selected Liquid Scintillation Cocktails. 0 ed. SCK CEN - Belgian Nuclear Research Center, 2008. 132 p. (SCK•CEN Reports; BLG-1052).

Bibtex - Download

@book{eba250f5928e49b9991909834d690ec0,
title = "A Performance Comparison of Nine Selected Liquid Scintillation Cocktails",
abstract = "In the selection of a suitable liquid scintillation (LSC) cocktail, the primary aspects taken into consideration are overall cocktail performance and specific laboratory needs. Overall performance of 9 selected, commercially available LSC cocktails was assessed by studying parameters of importance for the requirements of the Laboratory for Low Level Radioactivity Measurements of SCK•CEN: sample load capacity, sample compatibility, influence of sample load on counting efficiency, background count rate, figure of merit, quench resistance, sample stability and alpha/beta separation characteristics. The cocktails tested were EcoscintA, Insta Gel Plus, OptiPhase Hisafe3, OptiPhase Trisafe, Ready Gel, SafeScint 1:1, Ultima Gold, Ultima Gold LLT, and Ultima Gold XR. For the data acquisition a Packard TriCarb Model 1900CA and a Quantulus 1220 liquid scintillation counter is used. All samples were prepared in either 20 mL low potassium, borosilicate glass vials or 20 mL high density, polyethylene vials. The aim of this study was to determine a single cocktail that best suits all measurement requirements of the liquid scintillation laboratory at SCK•CEN for the determination of low levels of radioactivity in biological and environmental samples. As a conclusion, Optiphase HiSafe 3 was confirmed to be the optimal cocktail for the laboratory.",
keywords = "liquid scintillation cocktail performance comparison",
author = "Freddy Verrezen and Hilde Loots and Christian Hurtgen",
note = "RN - BLG-1052 Score = 2",
year = "2008",
month = "6",
language = "English",
volume = "1",
series = "SCK•CEN Reports",
publisher = "SCK CEN - Belgian Nuclear Research Center",
number = "BLG-1052",
address = "Belgium",
edition = "0",

}

RIS - Download

TY - BOOK

T1 - A Performance Comparison of Nine Selected Liquid Scintillation Cocktails

AU - Verrezen, Freddy

AU - Loots, Hilde

AU - Hurtgen, Christian

N1 - RN - BLG-1052 Score = 2

PY - 2008/6

Y1 - 2008/6

N2 - In the selection of a suitable liquid scintillation (LSC) cocktail, the primary aspects taken into consideration are overall cocktail performance and specific laboratory needs. Overall performance of 9 selected, commercially available LSC cocktails was assessed by studying parameters of importance for the requirements of the Laboratory for Low Level Radioactivity Measurements of SCK•CEN: sample load capacity, sample compatibility, influence of sample load on counting efficiency, background count rate, figure of merit, quench resistance, sample stability and alpha/beta separation characteristics. The cocktails tested were EcoscintA, Insta Gel Plus, OptiPhase Hisafe3, OptiPhase Trisafe, Ready Gel, SafeScint 1:1, Ultima Gold, Ultima Gold LLT, and Ultima Gold XR. For the data acquisition a Packard TriCarb Model 1900CA and a Quantulus 1220 liquid scintillation counter is used. All samples were prepared in either 20 mL low potassium, borosilicate glass vials or 20 mL high density, polyethylene vials. The aim of this study was to determine a single cocktail that best suits all measurement requirements of the liquid scintillation laboratory at SCK•CEN for the determination of low levels of radioactivity in biological and environmental samples. As a conclusion, Optiphase HiSafe 3 was confirmed to be the optimal cocktail for the laboratory.

AB - In the selection of a suitable liquid scintillation (LSC) cocktail, the primary aspects taken into consideration are overall cocktail performance and specific laboratory needs. Overall performance of 9 selected, commercially available LSC cocktails was assessed by studying parameters of importance for the requirements of the Laboratory for Low Level Radioactivity Measurements of SCK•CEN: sample load capacity, sample compatibility, influence of sample load on counting efficiency, background count rate, figure of merit, quench resistance, sample stability and alpha/beta separation characteristics. The cocktails tested were EcoscintA, Insta Gel Plus, OptiPhase Hisafe3, OptiPhase Trisafe, Ready Gel, SafeScint 1:1, Ultima Gold, Ultima Gold LLT, and Ultima Gold XR. For the data acquisition a Packard TriCarb Model 1900CA and a Quantulus 1220 liquid scintillation counter is used. All samples were prepared in either 20 mL low potassium, borosilicate glass vials or 20 mL high density, polyethylene vials. The aim of this study was to determine a single cocktail that best suits all measurement requirements of the liquid scintillation laboratory at SCK•CEN for the determination of low levels of radioactivity in biological and environmental samples. As a conclusion, Optiphase HiSafe 3 was confirmed to be the optimal cocktail for the laboratory.

KW - liquid scintillation cocktail performance comparison

UR - http://ecm.sckcen.be/OTCS/llisapi.dll/open/ezp_91988

UR - http://knowledgecentre.sckcen.be/so2/bibref/5223

M3 - BLG - Open report