A radon exhalation monitor

Research output: Contribution to journalArticle

Standard

A radon exhalation monitor. / Oberstedt, Stephan; Vanmarcke, Hans.

In: Radiation protection dosimetry, Vol. 63, No. 1, 1996, p. 69-72.

Research output: Contribution to journalArticle

Harvard

Oberstedt, S & Vanmarcke, H 1996, 'A radon exhalation monitor', Radiation protection dosimetry, vol. 63, no. 1, pp. 69-72. https://doi.org/10.1093/oxfordjournals.rpd.a031514

Author

Oberstedt, Stephan ; Vanmarcke, Hans. / A radon exhalation monitor. In: Radiation protection dosimetry. 1996 ; Vol. 63, No. 1. pp. 69-72.

Bibtex - Download

@article{047a85297a9141b7a689ab455684c03f,
title = "A radon exhalation monitor",
abstract = "A radon exhalation monitor has been constructed and calibrated. The monitor is based on the absorption of radon by active charcoal and subsequent measurements of its gamma activity by means of high resolution gamma spectrometry. Exhalation rates may be determined with an accurancy of better than 20{\%} within 48 h after exposure. The detection is estimated on the bais of counting statistics to about 0.1 mBq.m-2s-1.",
keywords = "radon, active charcoal, radon exhalation, exhalation monitor",
author = "Stephan Oberstedt and Hans Vanmarcke",
year = "1996",
doi = "10.1093/oxfordjournals.rpd.a031514",
language = "English",
volume = "63",
pages = "69--72",
journal = "Radioation Protection Dosimitry",
issn = "0144-8420",
publisher = "Oxford University Press",
number = "1",

}

RIS - Download

TY - JOUR

T1 - A radon exhalation monitor

AU - Oberstedt, Stephan

AU - Vanmarcke, Hans

PY - 1996

Y1 - 1996

N2 - A radon exhalation monitor has been constructed and calibrated. The monitor is based on the absorption of radon by active charcoal and subsequent measurements of its gamma activity by means of high resolution gamma spectrometry. Exhalation rates may be determined with an accurancy of better than 20% within 48 h after exposure. The detection is estimated on the bais of counting statistics to about 0.1 mBq.m-2s-1.

AB - A radon exhalation monitor has been constructed and calibrated. The monitor is based on the absorption of radon by active charcoal and subsequent measurements of its gamma activity by means of high resolution gamma spectrometry. Exhalation rates may be determined with an accurancy of better than 20% within 48 h after exposure. The detection is estimated on the bais of counting statistics to about 0.1 mBq.m-2s-1.

KW - radon

KW - active charcoal

KW - radon exhalation

KW - exhalation monitor

UR - https://ecm.sckcen.be/OTCS/llisapi.dll?func=ll&objId=31141548&objaction=overview&tab=1

U2 - 10.1093/oxfordjournals.rpd.a031514