Assessing 131I in thyroid by non-spectroscopic instruments - A European intercomparison exercise

Research output: Contribution to journalArticle

Authors

  • Mats Isaksson
  • David Broggio
  • Pavel Fojtík
  • Anne Laure Lebacq
  • Juan Francisco Navarro-Amaro
  • Jakub Osko
  • Begoña Pérez-López
  • Isabelle Vu
  • Paolo Battisti
  • Jimmy Börjesson
  • Marie Carlsson
  • Carlo-Maria Castellani
  • Magnus Gårdestig
  • Peter Hill
  • Grażyna Krajewska
  • Günter Lünendonk
  • Oliver Meisenberg
  • Mats Stenström
  • Sokratis El Mantani Ordoulidis

Institutes & Expert groups

  • CIEMAT - Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas
  • ENEA - Dipartimento Fusione e tecnologie per la Sicurezza Nucleare
  • Hallands sjukhus
  • Li.U - Linköping University
  • FZJ - Forschungszentrum Jülich GmbH
  • CLOR - Central laboratory for Radiological Protection
  • Falu Lasarett, Falun
  • Gävle sjukhus, Gävle, Sweden

Documents & links

DOI

Abstract

One of the issues of the Open Project for the European Radiation Research Area (OPERRA) was human thyroid monitoring in case of a large scale nuclear accident. This issue was covered in task 5.4 as project “CaThyMARA” (Child and Adult Thyroid Monitoring After Reactor Accident), which included several aspects of thyroid monitoring, e.g. screening of facilities able to perform thyroid monitoring in the European countries, dose estimation, modelling of detector response, and two intercomparison exercises. The intercomparison described in this paper focused on thyroid monitoring by non-spectrometric instruments, including gamma cameras and other instruments that were considered available for measurements made by members of the public. A total of 12 facilities from 7 European countries have participated and 43 various measuring devices have been evaluated. The main conclusion of this intercomparison is that the ability to make assessments of 131I activity in the thyroid to the exposed population after an accidental release must, on the average, be considered as good among the European laboratories taking part in this study. This intercomparison also gave the participants the possibility to calibrate the measuring devices for thyroid measurements of children where this procedure was not available before. A comprehensive report of the intercomparison is given.

Details

Original languageEnglish
Article number106115
Pages (from-to)1-6
Number of pages6
JournalRadiation Measurements
Volume128
DOIs
StatePublished - 1 Sep 2019
Event27th International Conference on Nuclear Tracks and Radiation Measurements - Strasbourg , France

Keywords

  • Thyroid, Iodine-131, Non-spectrometric, Measurements, Intercomparison

ID: 5704308