Efficiency calibration of BEGe and extended range detectors

Research output: Contribution to journalArticle

Standard

Efficiency calibration of BEGe and extended range detectors. / Bruggeman, Michel; Vidmar, Tim; Verheyen, Leen; Verrezen, Freddy (Peer reviewer).

In: Applied Radiation and Isotopes, Vol. 87, No. 5, 05.2014, p. 356-360.

Research output: Contribution to journalArticle

Author

Bibtex - Download

@article{4c06bd82bc8646048b6ecf7a3dcb0e2a,
title = "Efficiency calibration of BEGe and extended range detectors",
abstract = "Any approach to efficiency calibration of HPGe detectors sensitive to low-energy gamma-rays and X-rays has to deal with true coincidence summing. We developed a method based on the integration of the EFFTRAN code into the commercial Genie2K™ gamma-ray spectrometry software. EFFTRAN's role is to compute true coincidence summing corrections and to obtain satisfactory results for detectors sensitive to X-rays. EFFTRAN had to be upgraded to include a full simulation of the emission of X-rays. In this manner, efficiency calibrations were successfully set up for BeGe and XtRa detectors.",
keywords = "Gamma-ray spectrometry, True coincidence summing effects, Efficiency calibration",
author = "Michel Bruggeman and Tim Vidmar and Leen Verheyen and Freddy Verrezen",
note = "Score = 10",
year = "2014",
month = "5",
doi = "10.1016/j.apradiso.2013.11.051",
language = "English",
volume = "87",
pages = "356--360",
journal = "Applied Radiation and Isotopes",
issn = "0969-8043",
publisher = "Elsevier",
number = "5",

}

RIS - Download

TY - JOUR

T1 - Efficiency calibration of BEGe and extended range detectors

AU - Bruggeman, Michel

AU - Vidmar, Tim

AU - Verheyen, Leen

A2 - Verrezen, Freddy

N1 - Score = 10

PY - 2014/5

Y1 - 2014/5

N2 - Any approach to efficiency calibration of HPGe detectors sensitive to low-energy gamma-rays and X-rays has to deal with true coincidence summing. We developed a method based on the integration of the EFFTRAN code into the commercial Genie2K™ gamma-ray spectrometry software. EFFTRAN's role is to compute true coincidence summing corrections and to obtain satisfactory results for detectors sensitive to X-rays. EFFTRAN had to be upgraded to include a full simulation of the emission of X-rays. In this manner, efficiency calibrations were successfully set up for BeGe and XtRa detectors.

AB - Any approach to efficiency calibration of HPGe detectors sensitive to low-energy gamma-rays and X-rays has to deal with true coincidence summing. We developed a method based on the integration of the EFFTRAN code into the commercial Genie2K™ gamma-ray spectrometry software. EFFTRAN's role is to compute true coincidence summing corrections and to obtain satisfactory results for detectors sensitive to X-rays. EFFTRAN had to be upgraded to include a full simulation of the emission of X-rays. In this manner, efficiency calibrations were successfully set up for BeGe and XtRa detectors.

KW - Gamma-ray spectrometry

KW - True coincidence summing effects

KW - Efficiency calibration

UR - http://ecm.sckcen.be/OTCS/llisapi.dll/open/ezp_137782

UR - http://knowledgecentre.sckcen.be/so2/bibref/12068

U2 - 10.1016/j.apradiso.2013.11.051

DO - 10.1016/j.apradiso.2013.11.051

M3 - Article

VL - 87

SP - 356

EP - 360

JO - Applied Radiation and Isotopes

JF - Applied Radiation and Isotopes

SN - 0969-8043

IS - 5

ER -