Microdosimetric study at the CNAO active-scanning carbon-ion beam

Research output: Contribution to journalArticlepeer-review

Authors

Institutes & Expert groups

  • INFN - Instituto Nazionale di Fisica Nucleare
  • POLIMI - Politecnico di Milano
  • CNAO - National Centre of Oncological Hadrontherapy for the treatment of tumors

Documents & links

DOI

Abstract

The Italian National Centre for Oncological Hadrontherapy (CNAO) has been treating patients since 2011 with carbon-ion beams using the active-scanning modality. In such irradiation modality, the beam spot, which scans the treatment area, is characterised by very high particle-fluence rates (more than 105 s−1 mm−2). Moreover, the Bragg-peak is only ~1 mm-FWHM. Commercial tissue-equivalent proportional counters (TEPC), like the Far West Technologies LET-½, are large, hence they have limited capability to measure at high counting fluence rates. In this study we have used two home-made detectors, a mini-TEPC 0.81 mm2 in sensitive area and a silicon telescope 0.125mm2 in sensitive area, to perform microdosimetric measurements in the therapeutic carbon-ion beam of CNAO. A monoenergetic carbon-ion beam of 189.5 ± 0.3 MeV/u scanning a 3 × 3 cm2 area has been used. Spectral differences are visible in the low y-value region, but the mean microdosimetric values, measured with the two detectors, result to be pretty consistent, as well as the microdosimetric spectra in the high yvalue region.