Nanoscale calculation of the relative efficiency of 7LiF:Mg,Ti (MTS-7) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors for measuring electrons and positrons

Research output: Contribution to journalArticle

Standard

Bibtex - Download

@article{99d414bce07941bfbd5274f4d332e3c3,
title = "Nanoscale calculation of the relative efficiency of 7LiF:Mg,Ti (MTS-7) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors for measuring electrons and positrons",
abstract = "Using the Microdosimetric d(z) Model in combination with PHITS-simulated specific energy probability density distributions, the relative efficiency of 7LiF:Mg,Ti (MTS) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors was assessed as function of the incident energy for electrons and positrons spanning from 2 keV to 1 GeV. Additionally, the effect of the dopant concentration on the determined efficiency values was carefully investigated. Finally, the results are presented in combination with calculated specific energy frequency mean values and possible correlations were discussed.",
keywords = "Microdosimetry, Luminescent detectors, PHITS, Efficiency",
author = "Alessio Parisi and Lara Struelens and Filip Vanhavere",
note = "Score=10",
year = "2020",
month = "10",
day = "16",
doi = "10.1088/1742-6596/1662/1/012025",
language = "English",
volume = "1662",
pages = "1--4",
journal = "Journal of Physics: Conference Series",
issn = "1742-6588",
publisher = "IOP - IOP Publishing",

}

RIS - Download

TY - JOUR

T1 - Nanoscale calculation of the relative efficiency of 7LiF:Mg,Ti (MTS-7) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors for measuring electrons and positrons

AU - Parisi, Alessio

AU - Struelens, Lara

AU - Vanhavere, Filip

N1 - Score=10

PY - 2020/10/16

Y1 - 2020/10/16

N2 - Using the Microdosimetric d(z) Model in combination with PHITS-simulated specific energy probability density distributions, the relative efficiency of 7LiF:Mg,Ti (MTS) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors was assessed as function of the incident energy for electrons and positrons spanning from 2 keV to 1 GeV. Additionally, the effect of the dopant concentration on the determined efficiency values was carefully investigated. Finally, the results are presented in combination with calculated specific energy frequency mean values and possible correlations were discussed.

AB - Using the Microdosimetric d(z) Model in combination with PHITS-simulated specific energy probability density distributions, the relative efficiency of 7LiF:Mg,Ti (MTS) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors was assessed as function of the incident energy for electrons and positrons spanning from 2 keV to 1 GeV. Additionally, the effect of the dopant concentration on the determined efficiency values was carefully investigated. Finally, the results are presented in combination with calculated specific energy frequency mean values and possible correlations were discussed.

KW - Microdosimetry

KW - Luminescent detectors

KW - PHITS

KW - Efficiency

UR - https://ecm.sckcen.be/OTCS/llisapi.dll/open/40677056

U2 - 10.1088/1742-6596/1662/1/012025

DO - 10.1088/1742-6596/1662/1/012025

M3 - Article

VL - 1662

SP - 1

EP - 4

JO - Journal of Physics: Conference Series

JF - Journal of Physics: Conference Series

SN - 1742-6588

M1 - 012025

ER -