Neutron spectrum calibration using the Cd-ratio for multi-monitor method with a synthetic multi-element standard

Research output: Contribution to journalArticle

Authors

Institutes & Expert groups

Documents & links

Abstract

Several methods are in use for the determination of the thermal to epithermal neutron fluence rate ratio (f) and the deviation of the epithermal neutron spectrum from the 1/E shape parameter ( α). In our former work, it was proven that the recently developed and characterized Synthetic Multi-ELement Standard (SMELS) can be used for the fast verification of the stability of the irradiation parameters using the Au-Zr bare monitor method. However, this latter method using SMELS had a too low precision for an accurate determination of f and α. Therefore, the Cd-ratio for multi-monitor method using SMELS was investigated for two irradiation channels. As shown the material can also be used as a monitor for the calibration of an irradiation facility.

Details

Original languageEnglish
Pages (from-to)631-636
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume278
Issue number3
DOIs
Publication statusPublished - Jan 2009

Keywords

  • neutron activation analysis, neutron spectrum, Cd-ratio, multi-monitor method, SMELS

ID: 367256