On the Effectiveness of the Dynamic Force Adjustment for Reducing the Scatter of Instrumented Charpy Results

Research output: Report/bookBLG - Open reportpeer-review

Authors

Documents & links

Documents

Abstract

One of the key factors for obtaining reliable instrumented Charpy results is the calibration of the instrumented striker. An interesting alternative to the conventional static calibration recommended by the standards is the Dynamic Force Adjustment (DFA), in which forces and displacements are iteratively adjusted until equality is achieved between absorbed energies calculated under the test record (Wt) and measured by the machine encoder (KV). In this study, this procedure has been applied to the instrumented data obtained by 10 international laboratories using notched and precracked Charpy specimens, in the framework of a Coordinated Research Project (CRP8) of IAEA. DFA is extremely effective in reducing the between-laboratory scatter for both general yield and maximum forces. The effect is less significant for dynamic reference temperatures measured from precracked Charpy specimens using the Master Curve procedure, but a moderate reduction of the standard deviation is anyway observed. It is shown that striker calibration is a prominent contribution to the interlaboratory variability of instrumented impact forces, particularly in the case of maximum forces.

Details

Original languageEnglish
PublisherSCK CEN
Number of pages13
Edition0
Publication statusPublished - Sep 2008

Publication series

NameSCK•CEN Reports
PublisherStudiecentrum voor Kernenergie
No.BLG-1059

Keywords

  • Instrumented Charpy tests, striker calibration, dynamic force adjustment, IAEA CRP8, between-laboratory scatter, dynamic Master Curve

ID: 231281