Simulation of image detectors in radiology for determination of scatter-to-primary ratios using Monte Carlo radiation transport code MCNP/MCNPX

Research output: Contribution to journalArticle

Authors

Institutes & Expert groups

Documents & links

DOI

Abstract

The purpose of this study was to compare and validate three methods to simulate radiographic image detectors with the Monte Carlo software MCNP/MCNPX in a time efficient way. The first detector model was the standard semideterministic radiography tally, which has been used in previous image simulation studies. Next to the radiography tally two alternative stochastic detector models were developed: A perfect energy integrating detector and a detector based on the energy absorbed in the detector material. Validation of three image detector models was performed by comparing calculated scatter-to-primary ratios (SPRs) with the published and experimentally acquired SPR values. For mammographic applications, both the perfect energy integrating detector model and the blur-free energy absorbing detector model can be used to simulate image detectors, whereas for conventional x-ray imaging using higher energies, the blur-free energy absorbing detector model is the most appropriate image detector model. The radiography tally overestimates the scattered part and should therefore not be used to simulate radiographic image detectors.

Details

Original languageEnglish
Pages (from-to)2082-2091
JournalMedical Physics
Volume37
Issue number5
DOIs
Publication statusPublished - 22 Apr 2010

Keywords

  • Monte Carlo image detector simulation, scatter-to-primary ratio, radiography tally, MCNP/MCNPX