Some modifications to Sima's model for total efficiency calculation of well-type detectors

Research output: Contribution to journalArticlepeer-review

Authors

  • Stefaan Pommé
  • Johan Camps
  • Goedele Sibbens
  • Tim Vidmar
  • Yana Spasova

Institutes & Expert groups

Documents & links

Abstract

Some modifications are presented to Sima's simple analytical model for the calculation of the total detection efficiency of a well-type gamma-ray detector, aiming at improving its accuracy and broadening its applicability. The modifications pertain to some general improvements in the photon transmission probabilities through absorbing materials and implementation of specific equations for solid angle and mean path length through materials that describe the case of eccentrically positioned sources as well as volume sources. In particular, some improvement is achieved by implementing a contribution by elastic scattering effects.

Details

Original languageEnglish
Pages (from-to)143-147
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume281
Issue number1
DOIs
Publication statusPublished - 27 Jun 2009

Keywords

  • Detection efficiency, Well detector, Simulation, Analytical model, Standardisation

ID: 151088