Study of the dosimetric properties of Al2O3:C,Mg radiophotoluminescent (RPL) detectors: Master thesis report

Research output: ThesisMaster's thesis

Authors

  • Onur Murat

Institutes & Expert groups

Documents & links

Abstract

Aluminum oxide doped with carbon and magnesium (Al2O3:C,Mg ) was investigated to use together with immobilization masks used in radiotherapy treatments. In this way, thanks to RPL properties of the Al2O3:C,Mg the masks can be used for patient dosimetry after treatment sessions. Linearity, reproducibility, dose range and uniformity of the Al2O3:C,Mg material create challenges to use this powder in vivo dosimetry. The reading system is also studied to obtain higher sensitivity levels. Irradiated samples were read with Red and UV excitations. UV excitation with further optimized filter set up will be promising for future readings. The coating device can provide homogenous coating layer for samples. In 2D reading tests, reading protocol first optimized and height and surrounding effects were studied. In dose maps and spatial resolution tests the maximum resolution with certain aperture was seen. Increasing resolution with smaller aperture can create problems for sensitivity.

Details

Original languageEnglish
QualificationMaster of Science
Awarding Institution
  • ENSTA - École nationale supérieure de techniques avancées Paris
Supervisors/Advisors
Award date8 Sep 2015
Publication statusPublished - 8 Sep 2015

Keywords

  • radiophotoluminescence, dosimetry, immobilization mask

ID: 2152738