TEPC gas gain measurements in propane

Research output: Contribution to journalArticle

Authors

Institutes & Expert groups

Documents & links

DOI

Abstract

Knowledge of the gas gain is important to optimise the design and the operating characteristics of tissue-equivalent proportional counters (TEPCs), especially for simulated sites smaller than 1 µm. TEPC area monitors of the order of centimetres must operate at very low gas pressure to simulate micrometric volumes, consequently the Townsend theory cannot be applied: effects related to the presence of an electric-field gradient become important and must be considered. A detailed description of the electron avalanche formation is complex, but in most practical cases an analytical formula can be used. The so-called gradient-field model includes three characteristic constants of the counting gas, which were already experimentally determined for propane-tissue equivalent (TE) and dimethyl ether (DME) gases. The aim of this work is to measure the gas-dependent parameters for propane gas. Preliminary results obtained with a spherical TEPC are presented.

Details

Original languageEnglish
Pages (from-to)1-5
JournalRadiation protection dosimetry
Volume2014
DOIs
Publication statusPublished - 2 Feb 2014
EventNEUDOS 2013 -12th Symposium on Neutron and Ion Dosimetry - IRSN- Institut de Radioprotection et de Sûreté Nucléaire, Aix-en-provence, France
Duration: 3 Jun 20137 Jun 2013

Keywords

  • TEPC, microdosimetry, gas gain

ID: 114154