The impact of polyethylene vials on reactor channel characterization in k0-NAA

Research output: Contribution to journalArticle

Institutes & Expert groups

Documents & links

Abstract

Reactor channel characterization is commonly performed by irradiating bare and cadmium-covered fluence rate monitors, avoiding as much as possible the use of irradiation vials and spacers to position the monitors inside the channel. However, in routine k0-NAA often samples are packed in small polyethylene vials prior to irradiation. This work aims at studying the impact of typical polyethylene vials (1mm wall thickness) on the determination of the f and alpha channel specific parameters through the Bare, Cd-covered and Cd-ratio methods. The impact of these vials on each method was studied for 3 irradiation chanels of the BR1 reactor at SCK-CEN (Mol, Belgium) with low to high f and alpha values. The net impact was 1% on each parameter. Inconsistencies between the different methods were found when the impact of the polyethylene was neglected, implying that all determination methods must be pooled and thin monitors should be used for an accurate channel characterization.

Details

Original languageEnglish
Pages (from-to)569-575
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume286
Issue number2
DOIs
Publication statusPublished - 19 Aug 2010
Event16th Radiochemical Conference - Marianske Lazne, Czech Republic
Duration: 18 Apr 201023 Apr 2010

Keywords

  • Neutron Activation Analysis, f and alpha determination, k0-NAA

ID: 347485