Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project

Research output: Contribution to journalArticle

Standard

Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project. / Struelens, Lara; Carinou, Eleftheria; Clairand, Isabelle; Donadille, Laurent; Ginjaume, Mercè; Koukorava, Christina; Krim, Sabah; Mol, Harrie; Sans-Mercé, Marta; Vanhavere, Filip; D'Agostino, Emiliano (Peer reviewer).

In: Radiation Measurements, Vol. 46, No. 11, 11.2011, p. 1258-1261.

Research output: Contribution to journalArticle

Harvard

Struelens, L, Carinou, E, Clairand, I, Donadille, L, Ginjaume, M, Koukorava, C, Krim, S, Mol, H, Sans-Mercé, M, Vanhavere, F & D'Agostino, E 2011, 'Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project', Radiation Measurements, vol. 46, no. 11, pp. 1258-1261. https://doi.org/10.1016/j.radmeas.2011.08.002

APA

Vancouver

Struelens L, Carinou E, Clairand I, Donadille L, Ginjaume M, Koukorava C et al. Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project. Radiation Measurements. 2011 Nov;46(11):1258-1261. https://doi.org/10.1016/j.radmeas.2011.08.002

Author

Struelens, Lara ; Carinou, Eleftheria ; Clairand, Isabelle ; Donadille, Laurent ; Ginjaume, Mercè ; Koukorava, Christina ; Krim, Sabah ; Mol, Harrie ; Sans-Mercé, Marta ; Vanhavere, Filip ; D'Agostino, Emiliano. / Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project. In: Radiation Measurements. 2011 ; Vol. 46, No. 11. pp. 1258-1261.

Bibtex - Download

@article{5f6d2f6cca084d72945ffc9b0773818d,
title = "Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project",
abstract = "Although active personal dosemeters (APDs) are not used quite often in hospital environments, the possibility to assess the dose and/or dose rate in real time is particularly interesting in interventional radiology and cardiology (IR/IC) since operators can receive relatively high doses while standing close to the primary radiation field. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008e2011) supported by the European Commission within its 7th Framework Program. This paper reports on tests performed with APDs on phantoms using an X-ray facility in a hospital environment and APDs worn by interventionalists during routine practice in different European hospitals. The behaviour of the APDs is more satisfactory in hospitals than in laboratories with respect to the influence of the tube peak high voltage and pulse width, because the APDs are tested in scattered fields with dose equivalent rates generally lower than 1 Sv.h1.",
keywords = "Active personal dosemeters, interventional procedures, hospitals",
author = "Lara Struelens and Eleftheria Carinou and Isabelle Clairand and Laurent Donadille and Merc{\`e} Ginjaume and Christina Koukorava and Sabah Krim and Harrie Mol and Marta Sans-Merc{\'e} and Filip Vanhavere and Emiliano D'Agostino",
note = "Score = 10",
year = "2011",
month = "11",
doi = "10.1016/j.radmeas.2011.08.002",
language = "English",
volume = "46",
pages = "1258--1261",
journal = "Radiation Measurements",
issn = "1350-4487",
publisher = "Elsevier",
number = "11",

}

RIS - Download

TY - JOUR

T1 - Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project

AU - Struelens, Lara

AU - Carinou, Eleftheria

AU - Clairand, Isabelle

AU - Donadille, Laurent

AU - Ginjaume, Mercè

AU - Koukorava, Christina

AU - Krim, Sabah

AU - Mol, Harrie

AU - Sans-Mercé, Marta

AU - Vanhavere, Filip

A2 - D'Agostino, Emiliano

N1 - Score = 10

PY - 2011/11

Y1 - 2011/11

N2 - Although active personal dosemeters (APDs) are not used quite often in hospital environments, the possibility to assess the dose and/or dose rate in real time is particularly interesting in interventional radiology and cardiology (IR/IC) since operators can receive relatively high doses while standing close to the primary radiation field. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008e2011) supported by the European Commission within its 7th Framework Program. This paper reports on tests performed with APDs on phantoms using an X-ray facility in a hospital environment and APDs worn by interventionalists during routine practice in different European hospitals. The behaviour of the APDs is more satisfactory in hospitals than in laboratories with respect to the influence of the tube peak high voltage and pulse width, because the APDs are tested in scattered fields with dose equivalent rates generally lower than 1 Sv.h1.

AB - Although active personal dosemeters (APDs) are not used quite often in hospital environments, the possibility to assess the dose and/or dose rate in real time is particularly interesting in interventional radiology and cardiology (IR/IC) since operators can receive relatively high doses while standing close to the primary radiation field. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008e2011) supported by the European Commission within its 7th Framework Program. This paper reports on tests performed with APDs on phantoms using an X-ray facility in a hospital environment and APDs worn by interventionalists during routine practice in different European hospitals. The behaviour of the APDs is more satisfactory in hospitals than in laboratories with respect to the influence of the tube peak high voltage and pulse width, because the APDs are tested in scattered fields with dose equivalent rates generally lower than 1 Sv.h1.

KW - Active personal dosemeters

KW - interventional procedures

KW - hospitals

UR - http://ecm.sckcen.be/OTCS/llisapi.dll/open/ezp_117533