Validation of the determination of tin by k0-instrumental neutron activation analysis in foodstuff

Research output: Contribution to journalArticle

Standard

Validation of the determination of tin by k0-instrumental neutron activation analysis in foodstuff. / Vermaercke, Peter; Farina, Fulvio; Sneyers, Liesel; Bruggeman, Michel; Bouças, J.G.

In: Journal of Radioanalytical and Nuclear Chemistry, Vol. 281, No. 1, 06.2009, p. 35-39.

Research output: Contribution to journalArticle

Author

Bibtex - Download

@article{857cb516d6994d85bf820d46bb1f5180,
title = "Validation of the determination of tin by k0-instrumental neutron activation analysis in foodstuff",
abstract = "Several recent intercomparisons and certification exercises have shown that the determination of tin in e.g. food samples and plastics is not straightforward. k0-Instrumental Neutron Activation Analysis (k0-INAA) offers some intrinsic quality control opportunities for the determination of the tin content in samples since several monitoring radioisotopes are formed. In this work we will discuss the validation of the determination of tin by k0-INAA using different reference materials. The results show that only few of these radioisotopes are reliable in terms of accuracy and that from a survey of the tin content in a range of canned foods detection limits as low as 2 mg/kg can be achieved.",
keywords = "Tin, Neutron activation analysis, Foodstuff",
author = "Peter Vermaercke and Fulvio Farina and Liesel Sneyers and Michel Bruggeman and J.G. Bou{\cc}as",
note = "Score = 10",
year = "2009",
month = "6",
doi = "10.1007/s10967-009-0086-7",
language = "English",
volume = "281",
pages = "35--39",
journal = "Journal of Radioanalytical and Nuclear Chemistry",
issn = "0236-5731",
publisher = "Springer",
number = "1",

}

RIS - Download

TY - JOUR

T1 - Validation of the determination of tin by k0-instrumental neutron activation analysis in foodstuff

AU - Vermaercke, Peter

AU - Farina, Fulvio

AU - Sneyers, Liesel

AU - Bruggeman, Michel

AU - Bouças, J.G.

N1 - Score = 10

PY - 2009/6

Y1 - 2009/6

N2 - Several recent intercomparisons and certification exercises have shown that the determination of tin in e.g. food samples and plastics is not straightforward. k0-Instrumental Neutron Activation Analysis (k0-INAA) offers some intrinsic quality control opportunities for the determination of the tin content in samples since several monitoring radioisotopes are formed. In this work we will discuss the validation of the determination of tin by k0-INAA using different reference materials. The results show that only few of these radioisotopes are reliable in terms of accuracy and that from a survey of the tin content in a range of canned foods detection limits as low as 2 mg/kg can be achieved.

AB - Several recent intercomparisons and certification exercises have shown that the determination of tin in e.g. food samples and plastics is not straightforward. k0-Instrumental Neutron Activation Analysis (k0-INAA) offers some intrinsic quality control opportunities for the determination of the tin content in samples since several monitoring radioisotopes are formed. In this work we will discuss the validation of the determination of tin by k0-INAA using different reference materials. The results show that only few of these radioisotopes are reliable in terms of accuracy and that from a survey of the tin content in a range of canned foods detection limits as low as 2 mg/kg can be achieved.

KW - Tin

KW - Neutron activation analysis

KW - Foodstuff

UR - http://ecm.sckcen.be/OTCS/llisapi.dll/open/ezp_101299

UR - http://knowledgecentre.sckcen.be/so2/bibref/6242

U2 - 10.1007/s10967-009-0086-7

DO - 10.1007/s10967-009-0086-7

M3 - Article

VL - 281

SP - 35

EP - 39

JO - Journal of Radioanalytical and Nuclear Chemistry

JF - Journal of Radioanalytical and Nuclear Chemistry

SN - 0236-5731

IS - 1

ER -