Abstract
The objective of this study was to evaluate the feasibility of assessing the eye lens dose when lead glasses are used, by attaching a small radiophotoluminescent (RPL) glass dosemeter to the lead glasses. RPL dosemeters were placed over and under the shielding of the glasses and their response was evaluated for a range of parameters, including irradiations in a standard X-ray beam and Monte Carlo simulations of a cardiac/radiologic intervention (scattered beam). For the specific situations taken into account in this study, the optimal position was found to be a dosemeter placed close to the bridge of the lead glasses, over its shielding.
Details
Original language | English |
---|
Pages (from-to) | 257-261 |
---|
Number of pages | 5 |
---|
Journal | Radiation Measurements |
---|
Volume | 106 |
---|
Publication status | Published - 1 Nov 2017 |
---|
Event | SSD18 - 18th International Conference on Solid State Dosimetry - Marko Moscovitch School, Munich, Germany Duration: 3 Jul 2016 → 8 Jul 2016 |
---|